6 Ventilator Alarms
6.1
Setting Alarms
for alarm priority level, ranges, and descriptions for the Newport™
HT70 Ventilator.
The alarm controls are changed just like the parameter controls—with a simple touch/adjust/
accept method:
1.
Touch the
Alarms
button to enter the Alarms screen.
2.
Touch the desired alarm control (it will appear highlighted).
3.
Use the up and down arrow buttons to make the desired adjustment.
4.
Press the
Accept
button to confirm the change.
Several adjustments can be made before accepting the changes. When satisfied with the
changes, accept them all by pressing the
Accept
button once.
If the
Accept
button has not already been pressed, press
Cancel
to go back to the previous set-
tings.
The minute volume alarms are expiratory minute volume alarms when the on-airway flow
sensor is in use, and they are inspiratory minute volume alarms when the on-airway flow sensor
is not in use.
WARNING:
Setting any alarm limits to zero or off, or to extreme high or low values, can cause the associated
alarm not to activate during ventilation, which reduces its efficacy for monitoring the patient
and alerting the clinician to situations that may require intervention.
Note:
All configurable alarm settings are stored in the ventilator’s non-volatile internal memory, ensuring that
they are retained during power loss conditions or when the ventilator is powered off.
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......