Setup and Pre-Use Preparations
3-8
Operator's Manual
Figure 3-7.
Patient Circuit Connection to Humidifier
4.
Locate the 22 mm ID end of the main breathing circuit. Attach this end to the outlet port of the humid-
ifier chamber.
5.
Attach the prox inline filter with tubing to the prox line connector.
6.
Attach one end of the prox tubing to the prox inline filter.
7.
Attach the other end of the prox tubing to the prox line pressure port on the circuit tubing that is con-
nected to the inlet port of the humidifier chamber.
1
Breathing circuit
5
Prox line
2
Outlet port
6
Prox inline filter
3
Humidifier
7
Prox line connector
4
Pressure port
Summary of Contents for Newport HT70
Page 1: ...Operator s Manual Newport HT70 Ventilator with Accessories Plus Model TM...
Page 8: ...Page Left Intentionally Blank viii Operator s Manual...
Page 12: ...Page Left Intentionally Blank xii Operator s Manual...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Operator s Manual...
Page 72: ...Navigating the Screens Page Left Intentionally Blank 4 24 Operator s Manual...
Page 88: ...Operating the Ventilator Page Left Intentionally Blank 5 16 Operator s Manual...
Page 100: ...Ventilator Alarms Page Left Intentionally Blank 6 12 Operator s Manual...
Page 108: ...Battery Operation Page Left Intentionally Blank 7 8 Operator s Manual...
Page 138: ...Explanation of Modes and Controls Page Left Intentionally Blank 10 6 Operator s Manual...
Page 140: ...Quick Check Procedure Check Off Page Left Intentionally Blank A 2 Operator s Manual...
Page 142: ...Foldout Drawings B 2 Page left intentionally blank...
Page 144: ...Foldout Drawings B 4 Page left intentionally blank...
Page 146: ...Foldout Drawings B 6 Page left intentionally blank...
Page 147: ......