Software Upgrade Procedure
Service Manual
2-23
To remove and install the oxygen sensor, follow the procedure in section
2.9.3
Tubing
The tubing inside the ventilator does not need to be replaced at any predetermined time interval.
However, Covidien is aware that tubing may occasionally need replacing. During the overhaul
procedure, carefully inspect all tubing for degradation, cracks, or brittleness. If the tubing indicates
any of those symptoms, contact Customer Service to order the Newport™ e360 ventilator tube
replacement kit.
To ensure each tube is cut to the proper length and to replace the worn tubing in the ventilator,
refer to the system pneumatic diagram located in Appendix
2.9.4
Upon Completion
After an overhaul is completed, perform a complete electronic and pneumatic calibration as
described in Chapter
2.10
Software Upgrade Procedure
2.10.1
General Information
The following items are required for a software upgrade:
•
Latest software on USB flash drive
•
Software upgrade instructions
•
Reusable patient circuit (PBC340A) or equivalent
•
Cap to plug end of patient circuit (CAP100P) or equivalent
Note:
Perform the upgrade procedure to reload software after replacing a main board, display board, or the single-
board computer.
2.10.2
Upgrade Procedure
To upgrade the software, perform the following steps:
1.
After confirming the ventilator is off, connect the AC power cord to the AC wall outlet (see
for power entry module location).
2.
Press and hold the
Accept
button, and then turn on the power.
Summary of Contents for Newport e360
Page 1: ...Service Manual Newport e360 Ventilator with Accessories TM...
Page 7: ...v C Diagrams Index...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank viii...
Page 24: ...Introduction Page Left Intentionally Blank 1 12 Service Manual...
Page 56: ...Maintenance Overhaul and Software Upgrade Page Left Intentionally Blank 2 32 Service Manual...
Page 112: ...Calibration Procedures Page Left Intentionally Blank 5 16 Service Manual...
Page 146: ...Ordering and Contact Information Page Left Intentionally Blank B 6 Service Manual...
Page 148: ...Diagrams C 2 Service Manual Figure C 1 Newport e360 Ventilator Electrical Block Diagram...
Page 149: ...Service Manual C 3 Figure C 2 Newport e360 Ventilator Pneumatic Diagram SPD2100A...
Page 150: ...Diagrams Page Left Intentionally Blank C 4 Service Manual...
Page 152: ...Page Left Intentionally Blank 1 2...
Page 153: ......