Menu Options Navigation
Operator’s Manual
31
2.
PR Menu —
Provides access to pulse rate (PR) alarm limit settings and alarm inhi-
Figure 4-4.
QUICK ACCESS PR Menu with Alarm Audio OFF
To select alarm limit settings via Quick Access menus:
1.
Rotate the jog dial until the white highlight appears over the SpO
2
or the pulse
rate (PR) real-time value field.
2.
Press the jog dial.
3.
Rotate the jog dial until reaching the desired field.
•
Available SpO
2
alarm limit thresholds
–
SatSeconds™ alarm management values include OFF, 10, 25, 50, 100. The default
value is 100. Reference
SatSeconds™ Alarm Management Feature
–
Upper and lower SpO
2
alarm limit thresholds
–
SpO
2
alarm inhibition to disable audible alarms for SpO
2
limit violations
•
Pulse Rate alarm limits
–
Upper and lower pulse rate alarm limit thresholds
–
Pulse rate alarm inhibition to disable audible alarms for pulse rate limit vio-
lations
4.
Press the jog dial to select the field.