Adverse Events or Deviations
Operator’s Manual
105
Figure A-1.
Modified Bland-Altman Plot
A.5
Adverse Events or Deviations
The study was conducted as expected with no adverse events and no devia-
tions from the protocol.
1
Test Sensor;
Avg CO-oximeter value 70-100% SpO
2
2
Avg CO-oximeter value 70-100% SpO
2
Oximetry board with MAX-A sensor
Trendline of MAX-A sensor
Oximetry board with MAX-N sensor
Trendline of MAX-N sensor
Oximetry board with MAX-FAST sensor
Trendline of MAX-FAST sensor