Product Overview
2-4
Operator’s Manual
Display
Figure 2-2.
Display Components
8
LCD display panel
Use to view all graphic and numeric patient information as
well as status conditions and warning messages.
9
SpO
2
connector
Use to connect to the interface cable and SpO
2
sensor.
1
Upper and lower alarm limits
Reflects upper and lower SpO
2
and pulse rate alarm limits. An
alarm sounds each time patient oxygen saturation or pulse rate
values violate these alarm limits.
2
SpO
2
real-time value
Indicates hemoglobin oxygen saturation levels. Current upper
and lower alarm limit settings appear as smaller values to the
left of the dynamic SpO
2
value.
3
Time
Indicates the current time in hours, minutes, and seconds.
4
Pulse amplitude (blip bar)
Indicates pulse beat and the relative (non-normalized) pulse
amplitude. As the detected pulse becomes stronger, more bars
light with each pulse.
5
SatSeconds™ icon
Provides alarm management for mild or brief SpO
2
limit viola-
tions. When the SatSeconds™ feature is enabled, the SatSec-
onds™ icon fills in the clockwise direction as the SatSeconds™
alarm management system detects SpO
2
readings outside of
the limits setting. The SatSeconds™ icon empties in the counter-
clockwise direction when SpO
2
readings are within limits. When
the SatSeconds™ icon reaches full, a medium priority alarm
sounds. The adult default setting is 100. Reference
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......