Product Specifications
11-12
Operator’s Manual
Table 11-9.
Recommended Separation Distances
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the Monitoring System
(IEC/EN 60601-1-2)
The monitoring system is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of the monitoring system can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equip-
ment (transmitters) and the monitoring system as recommended below, according to the maximum output
power of the communications equipment.
Rated Maximum
Output Power (
P
) of
Transmitter in
Watts
Separation Distance According to Frequency of Transmitter in Meters
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.7 GHz
0.01
0.12
0.02
0.04
0.10
0.38
0.06
0.13
1.00
1.20
0.20
0.40
10.00
3.80
0.63
1.26
100.00
12.00
2.00
4.00
For transmitters rated at a maximum output power not listed above, the recommended separation distance
(
d
) in meters (
m
) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (
W
) according to the transmitter manufac-
turer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects, and people.
d
1.2
P
=
d
0.2
P
=
d
0.4
P
=
Summary of Contents for Nellcor Series
Page 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM...
Page 8: ...Page Left Intentionally Blank vi...
Page 12: ...Page Left Intentionally Blank x...
Page 20: ...Introduction Page Left Intentionally Blank 1 8 Operator s Manual...
Page 36: ...Installation Page Left Intentionally Blank 3 8 Operator s Manual...
Page 118: ...Product Specifications Page Left Intentionally Blank 11 18 Operator s Manual...
Page 156: ...Clinical Studies Page Left Intentionally Blank A 38 Operator s Manual...
Page 159: ......