Managing Oximetry Data
86
Operator’s Manual
Calculating the Analog Voltage Output
The OxiMax N-600x pulse oximeter data port provides analog voltage outputs
between pins 6, 13, 14, and ground (pin 10), which can be used to calibrate
oximeters such as a chart recorder. The voltage represents a specific measured
parameter's current value. The voltage differential varies proportionally from 0.0 to
+1.0 VDC as the pin's parameter varies over its full range of values. For example, as
the current value of SpO
2
varies from 0 to 100%, the voltage from pin 6 to ground
(pin 10) varies from 0.0 to +1.0 VDC. A voltage of 0.94 volts indicates a current SpO
2
value of 94.
To set analog voltage output
1. Press the SETUP softkey.
2. Press the NEXT softkey three times.
3. Press the ANALOG softkey.
Figure 85.
Analog Voltage Screen
a.
VOLT Options
—Selecting the 0 VOLT or 1 VOLT softkey causes that voltage to appear
at pins 6, 13, or 14 as referenced to ground pins 5 and 10.
b.
STEP Option
—Selecting the STEP softkey causes the voltage to increase from 0.0 to
1.0 volts at 1/10th-volt increments, with each step lasting at least one second.
4. Press the BACK softkey.
Have a qualified service technician perform the calibration of the attached device as
described in the
N-600x Service Manual
.
Table 13.
Analog Pinouts
Pin
Parameter
Parameter Range
6
Saturation
0 - 100%
13
Pulse Rate
0 - 250 bpm
14
Waveform
0 - 254 PAU
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......