Managing Oximetry Data
62
Operator’s Manual
Reading the Trend Data Display
Figure 46.
Identifying the components of the trend data display
View the trend display at any time. Selecting the one-hour trend display allows you
to view one hour of trend information. By using the scrolling feature, any single hour
of trend data can be viewed up to a maximum of 45 hours of trend information. The
ADJUST DOWN key below the ALARM SILENCE key scrolls the display to the left and
the ADJUST UP key below the ALARM SILENCE key scrolls the display to the right.
When the data displays, the most recent readings are on the right side of the graph.
The numbers below %SpO
2
indicate the highest and lowest parameter values at the
Table 8.
Trend Data Display Components
1
Highest and lowest reading at the cursor position.
2
Trend data that is being displayed (%SpO2, BPM, or PAU [pulse amplitude units]).
3
Amount of trend data displayed on the screen. Settings available are 20 and 40
seconds, 15 and 30 minutes, 1, 2, 4, 8, 12, 24, 36, and 48 hours.
4
No trend data recorded during this time.
5
Cursor can be moved left or right using the ADJUST UP (right) or ADJUST DOWN (left)
keys.
6
Time represented by the cursor (item 5).
7
Date represented by the cursor (item 5).
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......