Operator’s Manual
121
11 Product Specifications
Overview
This section contains OxiMax N-600x™ pulse oximeter physical and operational
specifications. Ensure all product requirements are met prior to installation of the
oximeter.
Physical Characteristics
Weight
5.8 lbs. (2.6 kg)
Dimensions
3.3 in. x 10.4 in. x 6.8 in. (8.4 cm x 26.4 cm x 17.3 cm)
Electrical Requirements
Power
Power Requirements
Rated at 100 to 120 volts AC (nominal 120 VAC) or
220 to 240 volts AC (nominal 230 VAC),
30 VA
Input Frequency
50/60 Hz
Fuses
Slow-blow 0.5 amp, 250 volts, IEC (5 x 20 mm)
Quantity: 2 external
Battery
Note:
The battery provides approximately seven hours of battery life when new and fully-charged
with no alarms, no serial data, no analog output, no nurse call output, with backlight on
while using a pulse simulator set for 200 bpm, high light and low modulation.
Type
Lead acid
Voltage
6 Volts DC
Recharge
8 hours with oximeter turned off
12 hours with oximeter turned on
Shelf Life
Four months, if oximeter runs on new, fully-charged battery
After four months storage, units run 33% of stated battery life
Compliance
91/157/EEC
Summary of Contents for Nellcor OxiMax N-600x
Page 1: ...Operator s Manual Nellcor N 600x Pulse Oximeter Alarm Management System OximaxTM Technology TM...
Page 10: ...Safety Information 4 Operator s Manual Page Left Intentionally Blank...
Page 21: ...Oximeter Features Operator s Manual 15 Figure 6 Second SpO2 Event No SatSeconds Alarm...
Page 26: ...General Introduction 20 Operator s Manual Page Left Intentionally Blank...
Page 126: ...Theory of Operations 120 Operator s Manual Page Left Intentionally Blank...
Page 145: ......