Troubleshooting
N-560 Service Manual
45
Table 6: Power Problems
Condition
Recommended Action
Battery Low indicator
lights steadily while
N
-
560 is connected to
AC and battery is fully
charged.
Ensure that the N-560 is plugged into an operational
AC outlet and the AC indicator is on.
Check the fuses. The fuses are located in the Power
Supply PCB as indicated in
on page 54. Replace if necessary.
Open the N-560 as described in
on page 54. Verify the power supply's
output to the battery while on AC. Disconnect the
battery leads from the battery and connect a DVM to
the battery leads. The voltage measured should be 13.8
VDC to 14.2 VDC and the current should be 500 mA to
600mA with the Power switch off and 750 mA to 830
mA with the Power switch on. Replace power supply if
above values are not met.
Check the cable connection between the Top Enclosure
and the Front Panel PCB. If the connection is good,
replace the Front Panel PCB. See
.
The N
-
560 does not
operate when
disconnected from AC
power.
The battery may be discharged. To recharge the battery,
refer to
on page 8. The N-560 may be
used with a less than fully charged battery, but with a
corresponding decrease in operating time from that
charge.
Replace the battery as indicated in
Battery Low indicator
on during DC operation
and an alarm is
sounding.
There are 15 minutes or less of usable charge left on the
N
-
560 battery before the
N-560
shuts off. At this point,
if possible, cease use of the N
-
560 on battery power,
connect it to an AC source and allow it to recharge
(approximately 6 hours). The N
-
560 may continue to
be used while it is recharging.
Summary of Contents for Nellcor OxiMax N-560
Page 1: ...Service Manual OxiMaxTM Technology Nellcor OxiMax N 560 Pulse Oximeter TM...
Page 10: ...List of Tables N 560 Service Manual viii...
Page 38: ...Performance Verification 28...
Page 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Page 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Page 84: ...Spare Parts 74...
Page 88: ...Packing for Shipment 78...
Page 104: ...Specifications 94...
Page 114: ...Data Port Interface Protocol 104...
Page 128: ...Technical Discussion 118...
Page 130: ...Technical Discussion 120...
Page 132: ...Technical Discussion 122...
Page 134: ...Technical Discussion 124...
Page 136: ...Technical Discussion 126...
Page 138: ...Technical Discussion 128...
Page 139: ...Technical Discussion N 560 Service Manual 129 Figure 55 Main PCB Parts Locator Diagram...
Page 140: ...Technical Discussion 130...
Page 142: ...Technical Discussion 132...
Page 144: ...Technical Discussion 134...
Page 146: ...Technical Discussion 136...
Page 147: ...Technical Discussion N 560 Service Manual 137 Figure 59 Front Panel Parts Locator Diagram...
Page 148: ...Technical Discussion 138...
Page 149: ...Technical Discussion N 560 Service Manual 139 Figure 60 Power Supply Schematic Diagram...
Page 150: ...Technical Discussion 140...
Page 151: ...Technical Discussion N 560 Service Manual 141 Figure 61 Power Supply Parts Locator Diagram...
Page 152: ...Technical Discussion 142...
Page 156: ......
Page 157: ......