System Features
Service Manual
3-11
The SatSeconds Safety Net
The SatSeconds “Safety Net” is for patients with saturation levels frequently
below the limit, but not staying below the limit long enough for the SatSec-
onds time setting to be reached. When three or more limit violations occur
within 60 seconds, an alarm sounds even if the SatSeconds time setting has
not been reached.
3.7.3
OxiMax SPD™ Alert Parameter
WARNING:
Supplemental oxygen will attenuate patterns of desaturation. A patient’s
respiratory compromise can be proportionally more severe before patterns
appear in the saturation trend. Remain vigilant when monitoring a patient on
supplemental oxygen.
Caution:
Do not modify any other alarm settings while using the SPD parameter.
The OxiMax SPD™ Alert (SPD) method of detecting patterns of desaturation in
adults is a function of the software within the monitoring system, which
detects repetitive occurrences of desaturation followed by resaturation. These
patterns are indicative of repetitive reductions in airflow through the upper
airway and into the lungs. With the SPD parameter enabled, the default value
for SatSeconds alarms is 100.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......