Return Authorization and Shipment
Service Manual
12-41
5.
Replace the Main PCB, ensuring all wires and cables remain clear. This is particu-
larly critical for the routing of the dog-leg flex cable and remaining flex cables
from the LCD display. Reference
Note:
The widest flex cable can be ordered separately. Ensure the label printed on the
cable, J1 DISPLAY SIDE, is visible before connecting to the J1 receptacle on the
LCD display.
6.
Replace the speaker, ensuring the speaker remains under the retaining clip on the
plate. Reference
7.
Reassemble the monitoring system chassis. Reference
12.8
Return Authorization and Shipment
12.8.1
General Instructions for Return
•
Contact Covidien Technical Services or a Covidien Service Center for a Returned
Goods Authorization (RGA) number and shipping instructions. Reference
•
If at all possible, pack in the original carton. If not, follow the directions for an
alternate carton.
•
Mark the carton and any shipping documents with the RGA number.
•
Select a method of shipment which provides tracking and proof of delivery.
•
Unless otherwise instructed, do not include any accessories, cables, or sensors.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......