Software and Firmware Upgrades
Service Manual
10-45
10.
Press the ENTER key on the keyboard.
11.
Press EXIT MENU.
10.9
Software and Firmware Upgrades
Covidien may provide platform software or firmware updates periodically.
Contact Covidien Technical Services to acquire or download the platform soft-
ware or firmware. Load platform software or firmware onto a virus-free USB
flash drive prior to proceeding with an upgrade.
Note:
Upgrades remove all trend data. If desired, export all trend data prior to upgrade.
Reference
To upgrade platform software or firmware
1.
Remove from monitoring.
2.
Access the service function menu. Reference
To access service function menus
3.
Select the SERVICE menu.
4.
Select one of the listed options, depending on the desired upgrade.
a.
SOFTWARE UPDATE
b.
FIRMWARE UPDATE
5.
Insert the USB flash drive containing the upgrade firmly into the USB port.
6.
Press NEXT, waiting for the upgrade to properly load. This may take time.
7.
Press FINISH.
8.
Remove the USB flash drive.
9.
Press EXIT MENU.
10.
Press EXIT AND RESTART.
11.
Press YES at EXIT PROGRAM prompt.
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......