Test Data Sheet Form
Service Manual
10-43
Table 10-7.
Electrical Safety Tests
Test Conditions
Limit (μA)
Results
Remarks
Earth leakage current (NC)
500
Pass / Fail
Value:
μA
Earth leakage current (SFC OS)
1000
Pass / Fail
Value:
μA
Earth leakage current (NCRM)
500
Pass / Fail
Value:
μA
Earth leakage current (SFC OSRM)
1000
Pass / Fail
Value:
μA
Enclosure leakage current (NC)
100
Pass / Fail
Value:
μA
Enclosure leakage current (OS)
500
Pass / Fail
Value:
μA
Enclosure leakage current (SFC OE)
500
Pass / Fail
Value:
μA
Enclosure leakage current (NCRM)
100
Pass / Fail
Value:
μA
Enclosure leakage current (SFC OSRM)
500
Pass / Fail
Value:
μA
Enclosure leakage current (SFC OERM)
500
Pass / Fail
Value:
μA
Patient leakage current (NC)
10
Pass / Fail
Value:
μA
Patient leakage current (OS)
50
Pass / Fail
Value:
μA
Patient leakage current (SFC OE)
50
Pass / Fail
Value:
μA
Patient leakage current (NCRM)
10
Pass / Fail
Value:
μA
Patient leakage current (SFC OSRM)
50
Pass / Fail
Value:
μA
Patient leakage current (SFC OERM)
50
Pass / Fail
Value:
μA
Mains voltage on applied part (SFC)
5000
Pass / Fail
Value:
μA
Mains voltage on applied part (SFCRM)
5000
Pass / Fail
Value:
μA
TESTS PERFORMED BY:
DATE:
NC
Normal Condition
OS
Single Fault Condition (Open Line/Neutral)
NCRM
Normal Condition Reverse
OSRM
Single Fault Condition (Open Line/Neutral) Reverse
SFC
Single Fault Condition
OE
Single Fault Condition (Open Earth)
OERM
Single Fault Condition (Open Earth) Reverse
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......