Calculating the Analog Voltage Output
Service Manual
7-17
7.5
Calculating the Analog Voltage Output
The data port provides analog voltage outputs between pins 6, 13, 14, and
ground (pin 10), which can be used to calibrate devices such as a chart record-
er. The voltage represents a specific measured parameter's current value. The
voltage differential varies proportionally from 0.0 to +1.0 VDC as the pin's
parameter varies over its full range of values. For example, as the current value
of SpO
2
varies from 0 to 100%, the voltage from pin 6 to ground (pin 10)
varies from 0.0 to +1.0 VDC. A voltage of 0.94 volts indicates a current SpO
2
value of 94.
Table 7-8.
Analog Pinouts
Pin
Parameter
Parameter Range
6
Saturation
0 - 100%
13
Pulse Rate
0 - 250 bpm
14
Waveform
0 - 254 PAU
Summary of Contents for Nellcor GR101704
Page 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Page 8: ...Page Left Intentionally Blank vi...
Page 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Page 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Page 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Page 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Page 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Page 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Page 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Page 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Page 235: ......