Clinical Studies
B-2
Operator's Manual
Results
:
Figure B-1.
rSO
2
and fSO
2
, 20 Subjects
One subject had a bias value of 37.2%, which corresponded with a very low signal quality index
number (SQI=1). The remaining 41 subjects had SQIs that started at 10 and remained greater than
four throughout the study.
Subject 206’s absolute bias data was excluded from the analysis due to the low signal quality, but
the trend data was left in as trending is not dependent on SQI.
Transitions 96 and 97 data; last 20 subjects; 179 data points. y=0.9133x–0.1755;
R
2
=0.9575; bias=0.26; standard deviation=2.9.
Table B-1.
Bias, Error, and Correlation for rSO
2
and fSO
2
Individual subjects (N=20)
mean
Range
Pooled (N=180) transi-
tions
Difference
1
(Bias)%
1.
Less subject 206
2.49
–6.47 to +11.78
0.26
Error (standard deviation)%
2.13
0.85 to 4.22
2.90
Correlation (r2)
0.947
0.823 to 0.990
0.958
Summary of Contents for INVOS 5100C
Page 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Page 12: ...Page Left Intentionally Blank x Operator s Manual...
Page 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Page 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Page 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Page 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Page 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Page 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Page 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Page 177: ......