Maintenance
9-6
Operator's Manual
9.7.3
Functional Test with Optional Field Test Device
Note:
Only a Covidien-supplied field test device should be used with the INVOS™ system. No other field test
devices should be connected to the system. For additional information on acquiring and using a field test
device, contact Covidien’s customer service department.
The field test device allows for functional testing of the INVOS™ system. To perform a functional
test with the field test device:
1.
Place the monitor on a sturdy, flat surface and turn on the monitor from the AC power source (mains)
(see
).
Note:
Use only hospital-grade AC mains outlets.
2.
Connect the preamplifier to the INVOS™ system monitor (see
).
3.
Connect the optional field test device to the preamplifier.
4.
Turn the mains power switch ON. Observe the illuminated blue
LED on the key panel.
5.
Turn the INVOS™ system on by pressing the ON/OFF
key to display the Start screen. Select
NEW
PATIENT
or
PREVIOUS PATIENT
to display the Main screen.
6.
On the Main screen, data from connected channels should appear. An rSO
2
of 50±1 should be dis-
played for all connected channels.
9.8
Troubleshooting Techniques
WARNING:
Electric shock hazard:
•
Before performing any service, turn off the unit, then disconnect the AC power (see
for location), remove the enclosure, and disconnect the battery from the DIB board.
•
Electrostatic discharge (ESD) protection also is required before performing any service on the
unit.
•
Make sure all cables are connected and all components are installed correctly before attempt-
ing to apply power or turn on the unit.
Summary of Contents for INVOS 5100C
Page 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Page 12: ...Page Left Intentionally Blank x Operator s Manual...
Page 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Page 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Page 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Page 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Page 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Page 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Page 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Page 177: ......