![Covidien Force FX-C User Manual Download Page 95](http://html1.mh-extra.com/html/covidien/force-fx-c/force-fx-c_user-manual_2673883095.webp)
Output Characteristics
Technical Specifications
Force FX Electrosurgical Generator C User’s Guide
9-11
Output Characteristics
Maximum Output for Bipolar and Monopolar Modes
Power readouts agree with actual power into rated load to within 15% or 5 watts,
whichever is greater.
* An indication of a waveform’s ability to coagulate bleeders without a cutting effect
Mode
Open Circuit
P–P Voltage
(max)
Rated Load
(max)
Power
(max)
Crest Factor*
Bipolar
Precise
Standard
Macro
450 V
320 V
750 V
100
Ω
100
Ω
100
Ω
70 W
70 W
70 W
1.5
1.5
1.5
Monopolar Cut
Low
Pure
Blend
1350 V
2300 V
3300 V
300
Ω
300
Ω
300
Ω
300 W
300 W
200 W
1.5
1.5
2.5
Monopolar
Coag
Desiccate
Fulgurate
LCF
Fulgurate
Spray
3500 V
8500 V
6900 V
9000 V
500
Ω
500
Ω
500
Ω
500
Ω
120 W
120 W
120 W
120 W
5
7.0
5.5
8