Preparing the Generator for Reuse
6-2
Force FX Electrosurgical Generator C User’s Guide
Preparing the Generator for Reuse
Step 1 – Disconnect the Accessories
A. Turn off the generator.
B. If applicable, remove the patient return electrode from the patient. Disconnect all
accessories from the front panel.
–
If the accessory is disposable (single use only), dispose of it according to the
procedures for your institution.
–
If the accessory is reusable, clean and sterilize it according to the manufacturer’s
instructions.
C. Disconnect and store any footswitch(es) used.
Step 2 – Clean the Generator
A. Turn off the generator, and unplug the power cord from the wall outlet.
B. Thoroughly wipe all surfaces of the generator and power cord with a mild cleaning
solution or disinfectant and a damp cloth.
Follow the procedures approved by your institution or use a validated infection control
procedure. Do not allow fluids to enter the chassis. The generator cannot be
sterilized.
Storing the Generator
If the generator is stored at a temperature outside its normal operating range of
50° to 104° F (10° to 40° C ), allow it to sit at room temperature for one hour prior to
use.
The generator can be stored indefinitely. However, if you store it longer than one year,
you must perform specific checkout procedures before use (refer to the service manual).
Caution
Do not reuse or resterilize accessories labeled “disposable” or “single use only.”
Warning
Electric Shock Hazard
—
Always turn off and unplug the generator before cleaning.
Notice
Do not clean the generator with abrasive cleaning or disinfectant compounds, solvents, or other
materials that could scratch the panels or damage the generator.