Preparing for Ultrasonic Electrosurgery
Befor
e Surgery
Force FX Electrosurgical Generator C User’s Guide
4-13
Preparing for Ultrasonic Electrosurgery
To prepare for ultrasonic electrosurgery, first prepare the CUSA System. Refer to the
CUSA System 200 User’s Guide
or
CUSA EXcel System User’s Guide
for assembly and
setup instructions.
Assemble and sterilize the CUSA handpiece and the CEM nosecone.
Connecting the Patient Return Electrode
Covidien recommends using a REM Polyhesive patient return electrode to maximize
patient safety. For further information, refer to
Applying a Patient Return Electrode to the
Connecting the CUSA Handpiece with CEM Nosecone
Warning
Electric Shock Hazard
—
• Do not connect wet accessories to the generator.
• Ensure that all accessories and adapters are correctly connected and that no metal is exposed.
Connect accessories to the proper receptacle. Improper connection of accessories may result in
inadvertent accessory activation or other potentially hazardous conditions. Follow the instructions
provided with electrosurgical accessories for proper connection and use.
Caution
Inspect accessories and cords (especially reusable accessories and cords) for breaks, cracks, nicks,
or other damage before every use. If damaged, do not use. Failure to observe this precaution may
result in injury or electrical shock to the patient or surgical team.