![Cosmo LS-R700 Operation Manual Download Page 97](http://html1.mh-extra.com/html/cosmo/ls-r700/ls-r700_operation-manual_2670610097.webp)
7 Operations Listed By Purposes
95
7
11.2 K(Ve) Check
Compares K(Ve) measured with a reference master tested part to the K(Ve) stored in memory.
This can be used for daily sensitivity checks of Type J models.
K(Ve) Check behavior will be same as the K(Ve) Automatic Setup.
K(Ve) Check Limit
Setting a tolerance in percentage (±) to the K(Ve) currently stored in memory.
> Main Menu > K(Ve) > K(Ve) Settings > Basic > K(Ve) Check Limit
Manual Operation
1) > Main Menu > K(Ve) > K(Ve) Check
2)
Press
►
to start K(Ve) Check.
Remote Operation
Transmit K(Ve) Check signal and Start Signals through control I/O port.
Results for K(Ve) Check
Lower than LL
Within the Limits
Larger than UL
DET LL
Pass
DET UL
When the Result was Fail
Perform K(Ve) Check after checking the followings and executing Mastering if the result was DET LL or
DET UL.
• Tested part
Check whether the tested part used for K(Ve) Check was same reference master part used for K(Ve)
Automatic Setup.
• Leak
Check the sealing surface for contaminants.
• If test results are not relatively consistent.
Normally extending Pressurization (CHG) timer or Equalization (BAL1) timer will help stabilize the
pressure and consequently the test result will be consistent.