92
7 Operations Listed By Purposes
7
Restore All
1)
Connect a USB memory to the USB port on the front panel.
2)
Execute Parameter Restore
> Main Menu > Settings > Backup/Restore > Restore > Restore All
“Initiating Test Parameter Restore OK to continue?” >
Yes
Individual Restore
In Individual Restore, test parameters set to the LS-R700 can be copied to
other LS-R700 testers. Items selected among the backup from the source
tester can be restored individually to the destination tester. However, this
feature requires full understanding that there are items that affect each other
such as timers, leak limits and K(Ve) value.
1)
Connect a USB memory to the USB port on the front panel.
2)
> Main Menu > Settings > Backup/Restore > Restore >
Individual Restore > Select parameters to be restored >
Enter
“Initiating Test Parameter Restore OK to continue?” >
Yes
8.2 Preparing for Replacing LS-R700
Perform System backup to prepare for restoring the test system in case of trouble in the future.
The current system settings can be restored easily to another testser from backup.
Backup
1) Insert a USB memory into the USB port on the front panel.
2) Execute the System Backup.
> Main Menu >System > System Backup/Restore > Backup >
“Initiating System Backup OK to continue?” >
Yes
Mastering Value, Compensation Value and Counter are
reset after Parameter Restore.
NOTE
NEVER
use this feature to copy the test parameters
to other testers.
The backup data can be restored only to itself.
Attention
Further individual items can be selected for
Advanced Se
.