8 Maintenance
8
When the Result was Fail
Perform K(Ve) Check after checking the followings and executing Mastering if the result was DET LL or
DET UL.
• Tested part
Check whether the tested part used for K(Ve) Check was same reference master part used for K(Ve)
Automatic Setup.
• Leak
Check the sealing surface for contaminants.
• If test results are not relatively consistent.
Normally extending Pressurization (CHG) timer or Equalization (BAL1) timer will help stabilize the
pressure and consequently the test result will be consistent.
4.2 No-Leak Check
No-Leak Check is a leak check of LS-R700 itself.
1) Close the both WORK- and MASTER-side stop valves on the rear panel.
2) > Main Menu > Maint. > Inspection > Leak Check > No-Leak Check
3) Check whether the displayed test pressure is appropriate.
Go to:
Charge Hold
4) Press
■
to cancel the Charge Hold.
5) If the test pressure is inappropriate, adjust it.
Go to:
Set Pressure
6) Press
►
to start a No-Leak Check.
7) Timers are fixed to the follows:
CHG=10.0s BAL1=0.5s BAL2=5.0s DET=10.0s
LS-R700 is not leaking if the result is within ±10 Pa.
If not, contact Cosmo for repair.
8) Press
■
to stop
9) Open the both WORK and MASTER side stop valves.
4.3 DPS Offset Adjustment
1)
> Main Menu > Maint. > Inspection > Sensor > DPS
2) Check whether the sensor is open to the atmosphere.
3) Select
Offset
.
4) Check whether the reading is within the tolerance.
Contact Cosmo for repair if the DPS reading exceeds the Offset Limit.