Service Manual Ponygraphic
Chapter 4
Maintenance
4.2
4. UNIT MAINTENANCE AND CONSUMABLE
REPLACEMENT
4.1.
Operator Maintenance Program
The following log is recommended to be maintained by the user or by operator on daily /Weekly
basis
General Precautions
•
DO NOT autoclave, gas sterilize or immerse the electrical parts of the in liquid.
•
The person operating the Pony graphic should not perform any servicing except as
specifically stated in the Pony graphic User's Manual.
•
If the Pony graphic fails to respond as described, DO NOT use it until you have
corrected the situation. Refer to the trouble-shooting section of this manual for more
detailed discussion.
•
Always turn the PONY graphic off before cleaning or servicing.
•
DO NOT immerse any electrical part (i.e. Optoelectronic transducer) in water or any
liquid.
•
DO NOT touch any exposed wiring or conductive surface while the cover is off and the
system is running if using a grounding wrist strap.
•
Use insulated tools when touching the components of the PCBs.
•
Electrostatic Discharge through the PCBs will damage the components of the system.
Handle all circuit boards (replacements and defectives) by their non conductive edges
and use anti-static containers when transporting them. Before servicing the equipment,
ground yourself and the tool to discharge any accumulated static charge by wearing a
static control wrist strap.
•
Check Power Supply Adapter voltage before connecting the AC main power.
•
Replace fuses ONLY with the same type and rating ones.
THINGS TO DO
DAILY
WEEKLY
MONTHLY
General
Clean & disinfect Turbine
√√√√
Charge batteries
√√√√
Clean the Unit
√√√√
(Follow Instructions in Users Manual for cleaning of the unit &Turbine.
Refer the Pony graphic User's Manual for cleaning instructions
Checks
Check
display
√√√√
Check Display backlight
√√√√
Check Cables (power, flow meter)
√√√√
Check battery status using the charger
√√√√
Calibration
Carry out Turbine calibration
√√√√
4.2. Replacement
Schedule
The replacement of parts should be followed for smooth functioning of the system.
THINGS TO DO
DAILY
MONTHLY
HALF YEAR YEARLY
Replace Internal Rechargeable batteries
(Every 2 years)
(For replacement of the batteries follow the service procedure section of this manual)
Refer Users manual for details on maintenance.
For any replacement of accessories please refer operators manual.
For any procedure for replacement of consumables refer section on Service procedures