Service Manual Ponygraphic
Chapter 3
Part Description
3.8
The flow and volume measurement is carried out as follows:
Volume
It is proportional to the number of pulses detected by the receiver diode during
expiration (expired volume) or inspiration (inspiratory volume);
Flow
the absolute value of the flow (positive for both expiration and inspiration) is
proportional to the inverse of the pulse width detected by the receiver diode;
IN/EX
The flow direction is detected by verifying which pulse (rx1, rx2) is received first. The
turbine doesn't need to be calibrated daily as it is not affected by pressure, humidity and
temperature.The turbine should be maintained with regular ordinary sterilization to move freely
without friction, free of dust.
3.6.3. Adjustment & checks
There are no adjustments to be performed on this assembly.
3.6.4. Assembly Drawing (Refer to Section: Schematics & Layouts)
3.6.5. Warnings & Indications in Event of Failure.
•
Flow and Volume errors
3.7.
Battery Charger (C00766-01-30) (C00767-01-30)
3.7.1. Function
The Battery charger of the Pony graphic supplies power to the Battery charging circuit (see
Mother Board).
Two models are available depending of the country where you purchased the system:
Pony graphic AC/DC adapter 220V 12V 1A
P/N: C00766-01-30
Pony graphic AC/DC adapter 110V 12V 1A
P/N: C00767-01-30
3.7.2. Description
The battery charger is a classified Class II equipment according to "IEC 601-1: 1990: Safety of
medical electrical equipment “General requirements"
The diagram in assembly drawing shows the polarity of the connections between the Pony
graphic Unit and the AC/DC adapter.
3.7.3. Adjustments
&
checks
There is no adjustment to be performed on this assembly
3.7.4. Assembly
Drawing