4 – Corning Lambda EliteTouch Pipettors 

INSTRUCTION MANUAL – 5

Model 

Volume 

Range 

(μL) 

Volume 

(μL)

Accuracy 

(%)

Permissible 

Volumes 

(μL)

Precision 

(%)

Fit to Tips 

Without 

Filter 

(μL)

Single-channel Pipettors

2

Min.

0.1

±40.0

0.06 – 0.14

≤12.0

10

0.2

±12.0

0.176 – 0.224

≤6.0

1

±2.7

0.973 – 1.027

≤1.3

Max.

2

±1.5

1.97 – 2.03

≤0.7

10

Min.

0.5

±4.0

0.48 – 0.52

≤2.8

10

1

±2.5

0.975 – 1.025

≤1.8

5

±1.5

4.925 – 5.075

≤0.7

Max.

10

±1.0

9.9 – 10.1

≤0.4

20

Min.

2

±3.5

1.93 – 2.07

≤1.5

200

10

±1.0

9.9 – 10.1

≤0.5

Max.

20

±0.8

19.84 – 20.16

≤0.3

50

Min.

5

±3.5

4.825 – 5.175

≤1.3

200

25

±1.0

24.75 – 25.25

≤0.5

Max.

50

±0.8

49.6 – 50.4

≤0.3

100

Min.

10

±3.0

9.7 – 10.3

≤1.0

200

50

±1.0

99.5 – 50.5

≤0.3

Max. 100

±0.8

99.2 – 100.8

≤0.2

200

Min.

20

±2.0

19.6 – 20.4

≤0.7

200

100

±1.0

99.0 – 101.0

≤0.3

Max. 200

±0.6

198.8 – 201.2

≤0.2

1000

Min. 100

±2.5

97.5 – 102.5

≤0.6

1000

500

±0.8

496 – 504

≤0.3

Max. 1000

±0.6

994 – 1006

≤0.2

8-channel and 12-channel Pipettors

8 - 10

12 - 10

Min.

0.5

±10.0

0.45 – 0.55

≤8.0

10

1

±8.0

0.92 – 1.08

≤6.0

5

±4.0

4.8 – 5.2

≤2.0

Max.

10

±2.0

9.8 – 10.2

≤1.2

8 - 50

12 - 50

Min.

5

±4.0

4.8 – 5.2

≤2.5

200

25

±3.0

24.25 – 25.75

≤1.2

Max.

50

±1.6

49.2 – 50.8

≤0.6

8 - 200

12 - 200

Min.

20

±3.0

19.4 – 20.6

≤1.5

200

100

±1.5

98.5 – 101.5

≤0.8

Max. 200

±1.0

198 – 202

≤0.6

8 - 300

12 - 300

Min.

30

±3.0

29.1 – 30.9

≤2.0

300

150

±1.2

148.2 – 151.8

≤1.0

Max. 300

±1.0

297 – 303

≤0.6

NOTE

•  The pipettor is designed for the transfer of liquids 

only in the tip. Do not draw liquids without the tip 

attached. The aspirated liquid should not get inside the 

pipettor, as it may damage the pipettor.

•  The single use of the tip reduces the risk of contamination 

of samples.

•  Keep the pipettor clean, avoiding the use of aggressive 

substances (e.g., acetone) for cleaning purposes.

•  Do not place the pipettor horizontally if there is liquid 

in the tip.

•  Only using the pipettor in accordance with the 

manufacturer’s instructions ensures the correct pipettor 

parameters are maintained.

•  After replacing the plunger or the shaft, the pipettor 

should be calibrated.

•  In the case of incorrect operation, the device should 

be cleaned in accordance with the instructions for use 

or transferred to a service point.

•  Ambient operating temperature: -5°C – 45°C 
•  Ambient storage conditions: -25°C – 55°C

 When working with the pipettor:

•  Follow general work safety regulations regarding hazards 

related to work in the laboratory.

•  Take special care when pipetting aggressive substances.
•  Use protective clothing, goggles, and protective gloves.
•  Avoid pointing the pipettor with the tip attached and 

filled at yourself or at others. 

•  Only use parts and accessories recommended by the 

manufacturer.

5. Specifications

The pipettor is a high quality instrument which offers 

excellent accuracy and precision. The accuracy and 

precision (repeatability) of the liquid volume depends on 

the quality of pipet tips used. The values for accuracy and 

precision shown in the table below were obtained using 

Corning® pipet tips. It is recommended to use these tips 

to obtain proper compatibility with the pipettors and proper 

parameters of accuracy and precision of the aspirated liquid.

Summary of Contents for Lambda EliteTouch

Page 1: ...Corning Lambda EliteTouch Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Page 2: ...6051 0 5 10 20 6052 2 20 50 6053 5 50 100 6054 10 100 200 6055 20 200 1000 6056 100 1000 8 channel and 12 channel Pipettors 8 10 12 10 6057 6061 0 5 10 8 50 12 50 6058 6062 5 50 8 200 12 200 6059 606...

Page 3: ...ector bar in the multi channel pipettors Materials Used Handle parts Polypropylene PP Polycarbonate PC Polyacetal POM Stainless steel Shaft and piston area parts Polypropylene PP Polyvinylidene fluori...

Page 4: ...e the pipettor as it may damage the pipettor The single use of the tip reduces the risk of contamination of samples Keep the pipettor clean avoiding the use of aggressive substances e g acetone for cl...

Page 5: ...ower position A B C The accuracy and precision are obtained gravimetrically using Corning pipet tips performing at least 10 measurements of distilled water at a temperature of 20 C 1 C according to th...

Page 6: ...unter operating mode switch the calibration switch located under the CAL sign on the counter overlay This switch should only be used when the pipettor is calibrated upper position marked as CAL on the...

Page 7: ...film remains relatively constant in successive pipetting operations with the same tip this error can be eliminated by forming the film before transferring the first sample This is done by aspirating a...

Page 8: ...holding the handle unscrew the remaining part of the shaft holder set from it Then proceed according to the pipettor models shown below See ISO 8655 for the full table of the value of the conversion f...

Page 9: ...d in the case of pipettor 2 L also the Seal P If you need to clean other components Unscrew the Shaft Cap M Remove from the Plunger I and pipetting Spring N Remove the Shaft J from the Connecting Bush...

Page 10: ...g Spring N from the shaft Remove the Shaft J from the Connecting Bush L If necessary remove the Seal P from the Plunger I Check the condition of components and clean as recommended Replace any damaged...

Page 11: ...ny damaged parts Assemble the components in reverse order When inserting the shaft or the shaft holder set into the Connecting Bush make sure that they are inserted all the way to the end by gently ro...

Page 12: ...ealing elements e g after repeated autoclaving procedures Sterilization Sterilization in an autoclave The pipettor can be sterilized in an autoclave at 121 C for 20 minutes Sterilization under other c...

Page 13: ...Shaft Kit with seal and O ring 2 SP9811 1 J O Shaft kit with O ring 10 SP9812 1 20 SP9813 1 J Shaft 50 100 SP9814 1 200 SP9815 1 1000 SP9816 1 L Connecting Bush All SP9810 1 H Tip ejector 2 10 SP9817...

Page 14: ...natural forces or other causes not arising from defects in the original material workmanship This warranty does not cover the PTFE seal O ring or shaft Claims for transit damage should be filed with t...

Page 15: ...EUROPE France t 0800 916 882 f 0800 918 636 Germany t 0800 101 1153 f 0800 101 2427 The Netherlands t 20 655 79 28 f 20 659 76 73 United Kingdom t 0800 376 8660 f 0800 279 1117 All Other European Coun...

Page 16: ......

Page 17: ...2019 Corning Incorporated All rights reserved 3 19 CLS AN 525 H EliteTouch IM Corning Incorporated Daniszewska 4 03 230 Warsaw www corning com lifesciences Made in Poland...

Reviews: