77
SERVICE PROCEDURE
Replacing cameras
When is necessary to replace a camera? It is necessary to replace a camera in the following cases:
-
During system calibration you receive message “Warning: bar is different” and the calibration pole
is ok.
-
On the calibration check a message “Bad calibration” come out.
-
Camera not recognized, hardware problem.
To check if the hardware work properly install ISCapture software.
Please insert the Aligner Software CD-ROM and on ADD-ON folder you will found the ISCapture 3.6 Setup.
Please Install it….
Click on Next.
Left camera
cover
Right camera
cover
GEO 15 Service Manual - Rel. 0
2
0
6
-2016