SENS:OFFS:PORT:DATA?
SENSe<Ch>:OFFSet:PORT<Pt>[:FREQuency]:DATA?
Description
Reads
out
the
array
of
the
frequency
points
of
port
<Pt>
when
the
frequency
offset
feature
is
ON
and
offset
type
is
"PORT"
(query
only).
The
array
size
is
N,
where
N
is
the
number
of
measurement
points.
For
the
n–th
point,
where
n
from
1
to
N:
<numeric
n>
the
frequency
value
at
the
n–th
point
Target
Port
<Pt>
of
channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
<Pt>={[1]|2}
Query
Response
<numeric
1>,
<numeric
2>,
…<numeric
N>
The
data
transfer
format
depends
on
the
FORM:DATA
command
setting.
Related
Commands
FORM:DATA
SENS:OFFS:STAT
SENS:OFFS:TYPE
Equivalent
Softkeys
None