Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
172
SENS:CORR:COLL:THRU
SENSe<Ch>:CORRection:COLLect[:ACQuire]:THRU
<rcvport>,<srcport>
Description
Measures
the
calibration
data
of
the
thru
standard
between
the
receiver
port
<rcvport>
and
the
source
port
<srcport>
(no
query).
Target
Channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Parameter
<rcvport>,
the
number
of
the
receiver
port
from
1
to
2
<srcport>,
the
number
of
the
source
port
from
1
to
2
Out
of
Range
If
an
incorrect
port
number
is
specified,
an
error
occurs
(error
code:
222).
If
the
same
port
numbers
are
specified,
an
error
occurs
(error
code:
220).
The
command
is
ignored.
Notes
The
command
start
the
measurement
for
the
channel
independently
of
the
trigger
and
trigger
source
settings.
The
command
waits
for
the
completion
of
the
measurement.
Equivalent
Softkeys
Calibration
>
Calibrate
>
Response
(Thru)
>
Thru
Calibration
>
Calibrate
>
One
Path
2–Port
Cal
>
Thru