Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
170
SENS:CORR:COLL:SAVE
SENSe<Ch>:CORRection:COLLect:SAVE
Description
Calculates
the
calibration
coefficients
from
the
calibration
standards
measurements
depending
on
the
selected
calibration
type.
On
completion
of
the
command,
all
the
calibration
standards
measurements
are
cleared
and
the
error
correction
automatically
turns
ON.
Before
executing
this
command
it
is
neccessery
to
select
calibration
type
by
one
of
commands
SENS:CORR:COLL:METH:XXXX
.
At
the
attempt
to
execute
this
command
before
all
the
needed
standards
are
measured,
an
error
occurs
and
the
command
is
ignored
(no
query).
Target
Channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Related
Commands
Calibration
type
selection:
SENS:CORR:COLL:METH:OPEN
SENS:CORR:COLL:METH:SHOR
SENS:CORR:COLL:METH:THRU
SENS:CORR:COLL:METH:ERES
SENS:CORR:COLL:METH:SOLT1
Calibration
standards
measurement:
SENS:CORR:COLL:LOAD
SENS:CORR:COLL:OPEN
SENS:CORR:COLL:SHOR
SENS:CORR:COLL:THRU
Equivalent
Softkeys
Calibration
>
Calibrate
>
{
Response
(Open)
|
Response
(Short)
|
Response
(Thru)
|
One
Path
2–Port
Cal
|
Full
1–Port
Cal
}
>
Apply