![Copper Mountain Technologies PLANAR TR1300/1 Programming Manual Download Page 63](http://html1.mh-extra.com/html/copper-mountain-technologies/planar-tr1300-1/planar-tr1300-1_programming-manual_2666189063.webp)
COM Server Commands
63
SCPI.CALCulate(
Ch
).SELected.LIMit.OFFSet.AMPLitude
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel
Ch
,
Ch
Description
The value of the limit line offset along Y–axis.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
0
Unit
dB (decibel) | ° (degree) | s (second)
Syntax
Dim Value As Double
Value =
app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.
AMPLitude
app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.
AMPLitude = –10
Equivalent
Softkeys
Analysis > Limit Test > Limit Line Offsets > Response Offset