COM Server Commands
52
SCPI.CALCulate(
Ch
).SELected.FUNCtion.DOMain.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
All traces of channel Ch (if the coupling is set to OFF by the
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle
property), the active trace of channel
Ch
(if otherwise),
Ch
Description
The ON/OFF state of the arbitrary range when executing the analysis by
the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute
method.
Allowable
Values
True
Arbitrary range ON
False
Arbitrary range OFF (entire sweep range)
Preset Value
False
Syntax
Dim Status As Boolean
Status =
app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.
STATe
app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.
STATe = True
Equivalent
Softkeys
None