51
INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
7 DIFFERENT HEATED SURFACES
The L100 Series Workstations have different heated surfaces.
Find the type of heated surfaces your workstation has, and find out which calibration value belongs to
which heat area.
Single T845/MP (3xregulation) right side
Type: T845/MP
Size: 800 x 430 mm
Size-MP 700 x 430
Regulation: 3
The safe area is located 50 mm from the
edge. In this case there are three different
temperature sensors that may need to be
calibrated. Each zone most be calibrated
separately, to calibrate "Zone 1", measure
only temperature in "Zone 1".
The red zone on the picture is "cross-heat"
zone; do not measure the temperature in
the cross-heat zone.
Single T845/MP (3xregulation) left side
Type: T845/MP
Size: 800 x 430 mm
Size-MP: 700 x 430
Regulation: 3
The safe area is located 50 mm from the
edge. In this case there are three different
temperature sensors that may need to be
calibrated. Each zone most be calibrated
separately, to calibrate "Zone 1", measure
only temperature in "Zone 1".
The red zone on the picture is "cross-heat"
zone; do not measure the temperature in
the cross-heat zone