INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
6.11.2 Calibrate Value Zone-1 (tn-1)
The tn-1 option is for the calibration of temperature on Zone-1 (See Section 7: Different Heated
Surfaces).
If there is an offset between the value shown on display and any measurements made with a high
precision external temperature sensor, this can be corrected. The new temperature reading will be
kept as the displayed value, with temperature control conducted on this basis. When the power is
disconnected default values will be restored (See Section 9: Temperature Calibration).
Follow these steps to navigate in the Calibrate Value Zone-1 (tn-1) option.
ACTION
KEY
DISPLAY
Press and hold both the UP
and
DOWN keys for 3 seconds
to enter the menu.
Press the DOWN key until the menu "StuP" appears in the
display.
Press the SP key once to enter the Setup menu.
Press the DOWN key until the option "tn-1" appears in the
display.
Press and hold the SP key to see the current value.
While holding the SP key, press either the UP
or
DOWN key
to select the calibration value.
When the required value has been selected, release the SP
key.
Press and hold both the UP
and
DOWN keys for 3 seconds
to exit the menu.
NOTE:
Always wait for the temperature to be stabilized before calibrating.
Summary of Contents for K-Systems L100 IVF Series
Page 1: ...K SYSTEMS IVF L100 Series Workstations Instruction Manual...
Page 76: ...76 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 77: ...77 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 78: ...78 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 79: ...79 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 80: ...80 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...