INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
6.10.4 UV Light On (u-on)
This function is only activated if a UV light is installed in the workstation. This option sets a start time for
the UV light.
Follow these steps to navigate in the UV Light On (u-on) option.
ACTION
KEY
DISPLAY
Press and hold both the UP and DOWN keys for 3 seconds to
enter the menu.
Press the DOWN key until the menu "- ti-" appears in the
display.
Press the SP key to enter the time menu.
Press the DOWN key until the option "u-on" appears in the
display.
To set the timer, press and hold the SP key.
Press the DOWN key to set the hour and the UP key to set the
minutes.
Press both the UP
and
DOWN keys to turn off the timer.
Press and hold both the UP and DOWN keys for 3 seconds to
exit the menu.
This function only works correctly if “ti.St” is set.
When the UV light is on, do not stand in front of the workstation.
NOTE:
To turn off the UV light press any key.
CAUTION
Never turn on the UV light when specimens are kept on the tabletop or
persons are near the Workstation.