Duette

®

M U LT I - B A N D   M U C O S E C TO M Y

cookmedical.com

M E D I C A L

QUICK REFERENCE GUIDE

Scope Compatibility

Prep 2

Insert the ligator handle into the endoscope 
accessory channel.  

Note:

 The irrigation adapter may be inserted 

through the white self-sealing valve prior to 
introducing the loading catheter. 

Prep 1

The firing position allows the handle to be 
rotated in the forward direction only. The  

two-way position

 allows the handle to rotate  

in both directions. Prior to introducing the 
endoscope, keep the handle in the two-way 
position.

Prep 4

Prep 5

Prep 3

Introduce either end of the loading catheter 
through the white seal in the ligator handle and 
advance, in short increments, until it exits the tip 
of the endoscope.

Attach the trigger cord, leaving approximately  
2 cm of cord between the knot and the hook. 
Withdraw the loading catheter and trigger cord 
up through the endoscope and out through the 
ligator handle.

Attach the barrel to the tip of the endoscope, 
advance as far as possible.

Olympus

Pentax

Fujinon

Order No.

Endoscope Outer Diameter

DT-6-5F

9.5 mm - 13 mm

DT-6

DT-6-XL

11 mm - 14 mm