6.1 Disconnecting the Power before Maintaining or Servicing the Chamber
To maintain or service this chamber, it is necessary to disconnect it from power.
1.
Turn the chamber power off by moving the toggle switch on the front panel to the
OFF position, as indicated by the symbol.
2.
Press and hold the UPS button until the power shuts off. Verify that all three LED
indicators are off.
3.
Turn off and lock the building disconnect switch provided during the installation.
4.
Remove the load plug from the UPS.
5.
Verify the power is off between contacts
L1A
and
NA
after opening a service access
panel as an additional precaution.
6.2 Cleaning the Chamber
For interior cleaning, dampen a clean towel or rag outside the unit, and carefully wipe the unit
down. Do not use abrasive cleaners. Mild detergents solutions are suitable for most cleaning
requirements.
Use glass cleaner on both the interior and the exterior of the glass viewport window, if present.
Refer to the
Conviron Chambers Maintenance & Troubleshooting Manual
for more information
on which cleaning solutions are appropriate for which chamber surface.
Warning: Burn hazard
Personal injury could result from contacting hot surfaces within the chamber.
The refrigeration system components and heater element become hot during normal
operation.
Allow the hot refrigeration components and heater elements to cool to the touch
before service.
Warning: Potential user injury after service
Personal injury to the users could result from not replacing the access panels after
service.
Ensure the access panel are replaced and secure before operating the chamber after
service.
Summary of Contents for GEN1000-GE
Page 1: ...GEN1000 GE OPERATOR MANUAL 291180 ENG R00...
Page 12: ...GEN1000 GE Operator Manual 4 November 8 2019 291180 ENG R00...
Page 16: ...GEN1000 GE Operator Manual 8 November 8 2019 291180 ENG R00...
Page 26: ...GEN1000 GE Operator Manual 18 November 8 2019 291180 ENG R00...
Page 34: ...GEN1000 GE Operator Manual 26 November 8 2019 291180 ENG R00...
Page 46: ...GEN1000 GE Operator Manual 38 November 8 2019 291180 ENG R00...
Page 53: ......