Instrument Exception Status
This register is available to verify the status of the instrument.
Instrument Control and I/O
This block of registers is available in some applications to give access to
monitor and/or control some of the instrument.
Register
Name
Comments
Read Only or
Read/Write
Type
41
Exception
Status
00 = no error
01 = analog input 1 failure
02 = analog input 2 failure
03 = analog input 3 failure
04 = analog input 4 failure
05 = invalid calibration parameter
06 = invalid reference parameter
07 = invalid property
08 to 09 reserved
10 = process parameters out of range
11 = input is over limit
12 = flow error detected
20 = system failure
21 = power supply is low
22 = new or failed clock battery
23 to 29 reserved
30 = alarm 1 active
31 = alarm 2 active
32 = alarm 3 active
33 = alarm 4 active
R
I
Register
Name
Comments
Read Only or
Read/Write
Type
42
Reserved
43
Logic Inputs
0 to 15 Binary representation of logic inputs
B0 = 0/1 (LSB)
input 1 activated/deactivated
B1 = 0/1
input 2 activated/deactivated
B2 = 0/1
input 3 activated/deactivated
B3 = 0/1
input 4 activated/deactivated
R
I
44
Operation Mode Representation of operation mode
0 = Idle/Local
Idle state
1 = Add
Addition calculation used
2 = Sub
Subtraction calculation used
R
I
45
Relay State
0 to 15 Binary representation of relay state.
0 = open; 1 = closed.
B0 = relay 1 (LSB)
B1 = relay 2
B2 = relay 3
B3 = relay 4
R
I