Contec CMS50D User Manual Download Page 1

   

 

 

Instructions to User 

Dear Users, thank you very much for purchasing our product. 
This  Manual  is  written  and  compiled  in  accordance  with  the

 

council  directive  MDD93/42/EEC  for  medical  devices  and 

harmonized standards. The Manual is written for the current Pulse Oximeter. In case of modifications and software upgrades, the 
information contained in this document is subject to change without notice. 
The  Manual  describes,  in  accordance  with  the  Pulse  Oximeter's  features  and  requirements,  main  structure,  functions, 
specifications,  correct  methods  for  transportation,  installation,  usage,  operation,  repair,  maintenance  and  storage,  etc.  as  well  as 
the safety procedures to protect both the user and equipment. Refer to the respective chapters for details. 
Please  read  the  Manual  very  carefully  before  using  this  equipment.  These  instructions  describe  the  operating  procedures  to  be 
followed  strictly,  failure  to  follow  these  instructions  can  cause  measuring  abnormality,  equipment  damage  and  personal  injury. 
The manufacturer is NOT responsible for the safety, reliability and performance issues and any monitoring abnormality, personal 
injury and equipment damage due to user's negligence of the operation instructions. The manufacturer’s warranty service does not 
cover such faults.   
Owing to the forthcoming renovation, the specific products you received may not be totally in accordance with the description of 
this User Manual. We would sincerely regret for that.   
This product is medical device, and can be used repeatedly. Its using life is 3 years.   

WARNING: 

 

The  uncomfortable  or  painful  feeling  may  appear  if  using  the  device  ceaselessly,  especially  for  the  microcirculation 
barrier patients. It is recommended that the sensor should not be applied to the same finger for over 2 hours. 

 

For the individual patients, there should be a more prudent inspecting in the placing process. The device can not be 
clipped on the edema and tender tissue. 

 

The light  (the infrared is invisible) emitted from the device is  harmful to the eyes, so the  user and the maintenance 
man, can not stare at the light. 

 

Testee can not use enamel or other makeup. 

 

Testee’s fingernail can not be too long. 

 

Please peruse the relative content about the clinical restrictions and caution. 

 

This device is not intended for treatment. 

Caution: Federal law restricts this device to sale by or on the order of a physician. 

 

1 Safety 

1.1 Instructions for Safe Operations 

 

Check the main unit and all accessories periodically to make sure that there is no visible damage that may affect patient's 
safety  and  monitoring  performance  about  cables  and  transducers.  It  is  recommended  that  the  device  should  be  inspected 
once a week at least. When there is obvious damage, stop using the monitor. 

 

Necessary maintenance must be performed by qualified service engineers ONLY. Users are not permitted to maintain it by 
themselves. 

 

The  oximeter  cannot  be  used  together  with  devices  not  specified  in  User's  Manual.Only  the  accessory  that  appointed  or 
recommendatory by manufacture can be used with this device. 

 

This product is calibrated before leaving factory. 

1.2 Warnings 

 

Explosive  hazard—DO  NOT  use  the  oximeter  in  environment  with  inflammable  gas  such  as  some  ignitable  anesthetic 
agents. 

 

DO NOT use the oximeter while the testee measured by MRI and CT. 

 

The person who is allergic to rubber can not use this device. 

 

The disposal of scrap instrument and  its accessories and packings(including battery, plastic bags, foams and paper boxes) 
should follow the local laws and regulations. 

 

Please check the packing before use to make sure the device and accessories are totally in accordance with the packing list, 
or else the device may have the possibility of working abnormally.   

 

Please don't measure this device with function test paper for the device's related information. 

1.3 Attentions 

 

Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and moisture. 

 

If the oximeter gets wet, please stop operating it. 

 

When it is carried from cold environment to warm or humid environment, please do not use it immediately.   

 

DO NOT operate keys on front panel with sharp materials. 

 

High temperature or high pressure steam disinfection of the oximeter is not permitted. Refer to User Manual in the relative 
chapter for instructions of cleaning and disinfection. 

 

Do not have the oximeter immerged in liquid. When it needs cleaning, please wipe its surface with  medical alcohol by soft 
material. Do not spray any liquid on the device directly. 

 

When cleaning the device with water, the temperature should be lower than 60

 

As to the fingers which are too thin or too cold, it would probably affect the normal measure of the patients' SpO

2

 and pulse 

rate, please clip the thick finger such as thumb and middle finger deeply enough into the probe. 

 

Do not use the device on infant or neonatal patients. 

 

The product is suitable for children above four years old and adults (Weight should be between 15kg to 110kg).   

 

The device may not work for all patients. If you are unable to achieve stable readings, discontinue use. 

 

The update period of data is less than 5 seconds, which is changeable according to different individual pulse rate. 

 

The  waveform  is  normalized.Please  read  the  measured  value  when  the  waveform  on  screen  is  equably  and  steady-going, 
Here this measured value is optimal value. And the waveform at the moment is the standard one. 

 

If some abnormal conditions appear on the screen during test process, pull out the finger and reinsert to restore normal use.  

 

The device has normal useful life for three years since the first electrified use.   

 

The hanging rope attached the product is made from Non- allergy material, if particular group are sensitive to the hanging 
rope, stop using it. In addition, pay attention to the use of the hanging rope , do not wear it around the neck avoiding cause 
harm to the patient. 

 

The  instrument  dose  not  have  low-voltage  alarm  function,  it  only  shows  the  low-voltage.please  change  the  battery  when 
the battery energy is used out. 

 

When the parameter is particularly, The instrument dose not have alarm function.Do not use the device in situations where 
alarms are required.   

 

Batteries must be removed if the device is going to be stored for more than one month, or else batteries may leak. 

 

A flexible circuit connects the two parts of the device. Do not twist or pull on the connection. 

1.4 Indication for Use   

The  Fingertip  Pulse  Oximeter  is  a  non-invasive  device  intended  for  the  spot-check  of  oxygen  saturation  of  arterial  hemoglobin 
(SpO

2

)  and  the  pulse  rate  of  adult  and  pediatric  patients  in  home  and  hospital  environments  (including  clinical  use  in 

internist/surgery, anesthesia, intensive care ect.). This device is not intended for continuous monitoring. 

  2 Overview 

The pulse oxygen saturation is the percentage of HbO

2

 in the total Hb in the blood, so-called the O

concentration in the blood. It 

is an important bio-parameter for the respiration. For the purpose of measuring the SpO

2

 more easily and accurately, our company 

developed the Pulse Oximeter. At the same time, the device can measure the pulse rate simultaneously. 
The  Pulse  Oximeter  features  in  small  volume,  low  power  consumption,  convenient  operation  and  being  portable.  It  is  only 
necessary for patient to put one of his fingers into a fingertip photoelectric sensor for diagnosis, and a display screen will directly 
show measured value of Hemoglobin Saturation. 

2.1 Classification 

Class II b, (MDD93/42/EEC    IX    Rule 10) 

2.2 Features 

 

Operation of the product is simple and convenient. 

 

The product is small in volume, light in weight (total weight is about 50g including batteries) and convenient in carrying. 

 

Power consumption of the product is low and the two originally equipped AAA batteries can be operated continuously for 
20 hours. 

 

The product will automatically be powered off when no signal is in the product within 5 seconds. 

 

4 directions display mode without waveform. 

 

Waveform display mode as line drawing or filling manner.   

2.3 Major Applications and Scope of Application 

The Pulse Oximeter can be used to measure human Hemoglobin Saturation and pulse  rate through finger, and indicate the pulse 
intensity by the bar-display. The product is suitable for use in  family, hospital (Ordinary sickroom), Oxygen Bar, social  medical 
organizations and also the measure of saturation oxygen and pulse rate. 

The product is not suitable for use in continuous supervision for patients.

 

The  problem  of  overrating  would  emerge  when  the  patient  is  suffering  from  toxicosis  which  caused  by  carbon 

monoxide, the device is not recommended to be used under this circumstance.

 

2.4 Environment Requirements 

Storage Environment 
  a) Temperature: -40

~+60

 

  b) Relative humidity: ≤95% 
  c) Atmospheric pressure: 500hPa~1060hPa 
Operating Environment 
  a) Temperature: 10

~40

 

  b) Relative Humidity: ≤75% 
  c) Atmospheric pressure: 700hPa~1060hPa   

 

3 Principle and Caution 

3.1 Principle of Measurement 

Principle  of  the  Oximeter  is  as  follows:  An  experience  formula  of  data  process  is  established  taking  use  of  Lambert  Beer  Law 
according  to  Spectrum  Absorption  Characteristics  of  Reductive  Hemoglobin  (Hb)  and  Oxyhemoglobin  (HbO

2

)  in  glow  & 

near-infrared zones. Operation principle of the instrument is: Photoelectric Oxyhemoglobin Inspection Technology is adopted in 
accordance with Capacity Pulse  Scanning  & Recording Technology, so that  two beams of different wavelength of  lights can be 
focused  onto  human  nail  tip  through  perspective  clamp  finger-type  sensor.  Then  measured  signal  can  be  obtained  by  a 
photosensitive element, information acquired through which will be shown on screen through treatment in electronic circuits and 
microprocessor. 

 

                                                                                    Figure 1 Operating principle 

3.2 Caution 

1.

 

The  finger should be placed properly (see the attached  illustration of this  manual, Figure 5), or else it  may cause  inaccurate 
measurement. 

2.

 

The SpO

2

 sensor and photoelectric receiving tube should be arranged in a way with the subject’s arteriole in a position there 

between

 

3.

 

The  SpO

2

  sensor  should  not  be  used  at  a  location  or  limb  tied  with  arterial  canal  or  blood  pressure  cuff  or  receiving 

intravenous injection. 

4.

 

Make sure the optical path is free from any optical obstacles like rubberized fabric. 

5.

 

Excessive ambient light  may affect the measuring result. It includes fluorescent lamp, dual ruby light, infrared heater, direct 
sunlight and etc. 

6.

 

Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy. 

7.

 

Testee can not use enamel or other makeup. 

3.3 Clinical Restrictions 

1.

 

As the  measure  is taken on the basis of arteriole pulse, substantial pulsating blood flow of subject  is required.  For a subject 
with weak pulse due to shock, low ambient/body temperature, major bleeding, or use of vascular contracting drug, the SpO

2

 

waveform (PLETH) will decrease. In this case, the measurement will be more sensitive to interference. 

2.

 

For those with a substantial amount of staining dilution drug (such as methylene blue, indigo green and acid indigo blue), or 
carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, and some with icterus problem, 
the SpO

2

 determination by this monitor may be inaccurate. 

3.

 

The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor blamed for serious error of 
SpO

2

 measure.   

4.

 

As the SpO

2

 value serves as a reference value  for judgement of anemic anoxia and toxic anoxia, some patients with serious 

anemia may also report good SpO

2

 measurement. 

 

4 Technical Specifications 

1)

 

Display Format:

 OLED Display; 

      SpO

2

 Measuring Range: 

0% ~ 100%; 

      Pulse Rate Measuring Range:

 30 bpm ~ 250 bpm; 

      Pulse Wave Display:

 columniation display and the waveform display. 

2)

 

Power  Requirements:

  2×1.5V  AAA  alkaline  battery  (or  using  the  rechargeable  battery  instead),    adaptable  range: 

2.6V~3.6V. 

3)

 

Power Consumption:

 Smaller than 30mA. 

4)

 

Resolution: 

1% for SpO

2

 and 1 bpm for Pulse Rate. 

5)

 

Measurement Accuracy:

 ±2%  in stage of 70%-100% SpO

2

, and  meaningless when  stage being smaller than 70%. ±2 bpm 

during the pulse rate range of 30-99 bpm and ±2% during the pulse rate range of 100~250 bpm .   

6)

 

Measurement  Performance  in  Weak  Filling  Condition:

  SpO

2

  and  pulse  rate  can  be  shown  correctly  when  pulse-filling 

ratio is 0.4%. SpO

2

 error is ±4%, pulse rate error is ± 2 bpm during the pulse rate range of 30~99 bpm and ±2% during the 

pulse rate range of 100~250 bpm . 

7)

 

Resistance  to  surrounding  light:

  The  deviation  between  the  value  measured  in  the  condition  of  man-made  light  or  indoor 

natural light and that of darkroom is less than ±1%. 

8)

 

It is equipped with a function switch. The Oximeter can be powered off in case no finger is the Oximeter within 5 seconds.     

9)

 

Optical

 

Sensor

 

      Red light (wavelength is 660nm, 6.65mW)   
      Infrared    (wavelength is 880nm, 6.75mW)                                                                     

  5 Accessories 

 

One hanging rope; 

 

Two batteries(optional) 

 

One User Manual. 

  6 Installation 

6.1 View of the Front Panel 

         

 

                          Figure 2 Front view                        Figure 3 Batteries installation                    Figure 4 Mounting the hanging rope 

6.2 Battery 

Step 1. Refer to Figure 3. and insert the two AAA size batteries properly in the right direction. 
Step 2. Replace the cover.   

Please take care when you insert the batteries for the improper insertion may damage the device. 

 

6.3 Mounting the Hanging Rope   

Step 1. Put the end of the rope through the hole. 
Step 2. Put another end of the rope through the first one and then tighten it. 

 

7 Operating Guide 

1)

 

Insert the two batteries properly to the direction, and then replace the cover. 

2)

 

Open the clip as shown in Figure 5.   

 

                                                                      Figure 5 Put finger in position 
3)

 

Let the patient’s finger put into the rubber cushions of the clip (make sure the finger is in the right position),      and then clip 
the finger. 

4)

 

Press the switch button once on front panel. 

5)

 

Do  not  shake  the  finger  and  keep  the  patient  at  ease  during  the  process.  Meanwhile,  human  body  is  not  recommended  in 
movement status. 

6)

 

Get the information directly from screen display. 

7)

 

The button

has three functions.When the device is power off, pressing the button can open it; When the device is 

power on, pressing the button shortly can change direction of the screen; When the device is power on, pressing the button 

long can change brightness of the screen. 

        Press the button shortly when the device is power on, the display mode will change, show as follow figure: 