background image

EVS1000 Blot units 

 

3

 

 

Electroblotting 

Setting up the blot sandwich 

 

Each blot sandwich should be set up as follows: 

 

Cassette clamp [-ve] (black) side placed in a tray or other suitable surface. 

 

Pre-soaked fibre pad (note two can be used with thin gels). 

 

Two pieces of thick filter paper, about 2...3 mm thick, pre-soaked in buffer. 

 

Gel. 

 

Transfer membrane. Usually this requires pre-soaking but consult the 

manufacturer’s

 

instructions for the type of membrane you are using. This should be smoothed so that no air 
bubbles have been trapped. 

 

Two pieces of thick filter paper, about 2...3 mm thick, pre-soaked in buffer. 

 

Pre-soaked fibre pad (note two can be used with thin gels). 

 

Cassette clamp [+ve] (red) side slotted into the groove in the bottom of the black cassette. 

 

Close the hinge carefully so as to not disturb the sandwich. 

 

Fill the tank with buffer solution up to the maximum fill line indicated on the side of each unit. 
Improved transfer can usually be obtained by using chilled buffer. 

Buffer Volume 

EVS1100 

EVS1200 

EV1300 

One cassette

 

1380 ml 

2800 ml 

5600 ml 

Two cassettes

 

1290 ml 

2620 ml 

5240 ml 

Three cassettes

 

1200 ml 

2440 ml 

4880 ml 

Four cassettes 

1110 ml 

2260 ml 

4520 ml 

 

Each cooling pack will take the place of 100 ml of buffer for EVS1100 and 500 ml of buffer for 
EVS1200 and EVS1300. 

Blot run conditions 

 

Insert the cassettes into the slots in the module with the black side of each adjacent to the negative 
electrode. It is a good idea to note the orientation and order the blot sandwiches were loaded in. This 
can be done by noting which samples were loaded adjacent to each electrode. 

 

Use of a magnetic stirring bar and plate is recommended to mix the buffer to give consistency of 
transfer. A 4 mm diameter stirring bar should be placed underneath the module, in the centre of the 
tank. The cooling pack provided, pre-frozen, can be inserted at the side or front of the tank for 
extended blots. Additional cooling packs can be purchased as accessories to further aid cooling. 

 

Insert the module, fit the lid and connect to a power supply. 

 

Consult below table for details on recommended power supply voltage settings and blot times. Please 
note voltage and current will vary according to the amount of cassettes, type and temperature of 

Summary of Contents for EV1300

Page 1: ...EVS1000 BLOT series MANUAL Rev 44 2019 ...

Page 2: ......

Page 3: ...tenance 1 Environmental conditions 2 Fitting electrode cables 2 Electroblotting 3 Setting up the blot sandwich 3 Blot run conditions 3 Buffer solutions for blotting 4 Protein Western blotting buffers 5 Nucleic acid transfer buffers 5 Transfer buffers 5 Certificate 6 Warranty 7 ...

Page 4: ...s of various gases are produced at the electrodes The type of gas produced depends on the composition of the buffer employed To disperse these gases make sure that the apparatus is run in a well ventilated area General care and maintenance Units are best cleaned using warm water and a mild detergent Water at temperatures above 60 C can cause damage to the unit and components The inner module shoul...

Page 5: ...tion occurs Occasionally however a temporary conductivity caused by condensation must be expected Fitting electrode cables Note Before setting up the tank please ensure that it has been properly cleaned and dried Note the position of the lid on the unit This shows the correct polarity and the correct orientation of the cables black is negative and red positive Remove the lid from the unit Note If ...

Page 6: ...assette 1380 ml 2800 ml 5600 ml Two cassettes 1290 ml 2620 ml 5240 ml Three cassettes 1200 ml 2440 ml 4880 ml Four cassettes 1110 ml 2260 ml 4520 ml Each cooling pack will take the place of 100 ml of buffer for EVS1100 and 500 ml of buffer for EVS1200 and EVS1300 Blot run conditions Insert the cassettes into the slots in the module with the black side of each adjacent to the negative electrode It ...

Page 7: ...mbrane from the gel The membrane is now ready to be probed Duration of blotting EVS1100 EVS1200 EV1300 One hour 100 V 400 mA 100 V 400 mA 100 V 400 mA Three hours 50 V 200 mA 50 V 200 mA 50 V 200 mA Buffer solutions for blotting Do not adjust the pH when making these buffers as this will cause blot over heating The pH will vary according to the freshness of the reagents used Towbin Buffer 25 mM Tr...

Page 8: ...orthern Blotting Buffer For high efficiency transfer of RNA from agarose gels Soak 12 pieces of standard grade filter paper 6 for the cathode 6 for the anode in o 50 x 0 2 M morpholinopropanesulfonic acid MOPS o 50 mM sodium acetate o 5 mM EDTA pH 7 0 Nucleic acid transfer buffers Soak 12 pieces of standard grade filter paper 6 for the cathode 6 for the anode in 1 x TAE 40 mM Tris pH 7 6 20 mM ace...

Page 9: ...EVS1000 Blot units 6 Certificate ...

Page 10: ...ce free of charge units where improper solutions or chemicals have been used For a list of these please see the Care and Maintenance subsection Consort products are for research use only Consort is not liable for consequential damages arising out of the use or handling of its products A return authorisation must be obtained from Consort before returning any product for warranty repair on a freight...

Page 11: ...EVS1000 Blot units 8 ...

Page 12: ...EVS1000 Blot units 9 Consort bvba Hertenstraat 56 unit 9 2300 Turnhout Belgium Tel 32 0 14 41 12 79 Fax 32 0 14 42 91 79 Sales sales consort be Support support consort be Information info consort be ...

Reviews: