18
EMAILING
The IWS has an advanced feature that allows emailing in the event of an alarm or an event. The IWS is
only capable of sending email. It cannot receive email. PROPER SETUP OF THE EMAIL FUNCTION
FOR THE IWS REQUIRES A KNOWLEDGEABLE LAN ADMINISTRATOR.
EMAIL1
EMAIL1 is used to report “alarm” conditions only. Reported alarm conditions are:
Error
– “description”
Left Side O/P
– “value” (O/P = over pressure)
Right Side O/P
– “value”
Reserve Alarm
Left Side U/P
– “value” (U/P = under pressure)
Right Side U/P
– “value”
Reserve In-Use
Delivery Regulator O/P
– “value”
Delivery Regulator U/P
– “value”
Delivery Regulator Status
– “value”
Alarm Clear U/P Left Side
– “value”
Alarm Clear O/P Left Side
– “value”
Alarm Clear U/P Right Side
– “value”
Alarm Clear O/P Right Side
– “value”
Alarm Clear Reserve Gas
The word “value” is replaced by a pressure in psi, and the word “description” is replaced with an
explanation of what happened when an email is sent.
To access the EMAIL1 screen, mouse click on the “EMAIL1” button on the EDIT screen. An image
similar to Figure 16 should appear on your screen.
Table 3 provides an explanation of the contents of the EMAIL1 screen