Concoa MEDICAL INTELLISWITCH 571 Series Manual Download Page 13

 

13 

IntelliSwitch.  This gives the user an opportunity to review and/or modify any value before the parameters 
are permanently changed.  The other buttons will be explained further along in this manual. 
 
Pop-up messages are used throughout this screen to aid in setting parameters.  Items 1-10 and buttons 
“ZERO SET”, and “RESTORE DEFAULTS” have associated pop-up messages that will appear when the 
mouse pointer is over them.  The “SET” button and “UNDO” buttons will show pop-up messages after 
they have been selected.  Use these messages as guides to correctly set values on this screen. 
 
Figure 7 and Table 2 describe the various parameters on the screen. 

 

Table 2 

 

Bullet 

Item 

Description 

Switchover Pressure 

This is a preset pressure in the IntelliSwitch used to determine 
when an inlet can no longer provide enough gas pressure to the 
delivery regulator.  Data range: 50-500psi 

Look Back 

This is a preset time in minutes used for liquid cylinders to 
determine when to look back at an inlet pressure to determine if 
pressure has been restored above the switchover point. The 
IntelliSwitch will look back at an inlet after the look back time has 
expired to see if its pressure has built back up above the 
switchover point.  If it has, then the IntelliSwitch will switch back to 
that side to try to use more of the gas.  Data range: 10-60 min. 

Switch back 

A preset time in minutes used for liquid cylinders in conjunction 
with the look back time.  If the look back feature is invoked and 
after time out has determine that it can switch back to a side 
whose pressure has built back, a switch back timer will start.  If the 
inlet pressure on the same side drops again before the switchback 
time has expired the IntelliSwitch determines that that side is truly 
empty and will alarm and switch to the other inlet.  Data range: 10-
60 min. 

Hysteresis 

This is a preset pressure value that is used to prevent the 
IntelliSwitch from “dithering” at the switchover point.  Data range: 
10-50 psi. 

Left Offset 

This is a value that can be used to adjust the pressure transducer 
reading for the left inlet.  This value should only be changed if it is 
necessary to recalibrate the transducer.  Changing this value will 
change the left inlet pressure reading.  A (+) or (-) value may be 
entered here.  Refer to “Calibrating Transducers” section of the 
IntelliSwitch user

’s manual.  Data range: +/- 200 psi. 

Right Offset 

This is a value that can be used to adjust the pressure transducer 
reading for the right inlet.  This value should only be changed if it 
is necessary to recalibrate the transducer.  Changing this value 
will change the right inlet pressure reading.  A (+) or (-) value may 
be entered here.  Refer to “Calibrating Transducers” section of the 
IntelliSwitch users manual.  Data range: +/- 200 psi. 

System ID 

This is a number that can be assigned by the user to identify the 
IntelliSwitch.  Data range:  1-254. 

GAS Type 

Displays the gas type the IntelliSwitch is controlling.  The pop-up 
message will show the choices.  The choices for gas type are 
numbered.  Enter the number next to the gas type in the text box.  
The name of the gas will appear on the status screen next to units 
of measure   Data range:1-99 (refer to pop-up message for details) 

Delivery Pressure 
Tolerance 

This is the variation in delivery regulator pressure that is allowed 
before the IntelliSwitch switches regulators and alarms.  The value 
is +/- the pressure shown.  Data range: 5-15psi. 

Summary of Contents for MEDICAL INTELLISWITCH 571 Series

Page 1: ...ION AND USER S GUIDE Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation of America 1501 Harpers Road Virginia Beach VA 23454 Telephone 1 800 225 0473 or 757 4...

Page 2: ......

Page 3: ...ing a Parameter 15 EMAILING 18 EMAIL1 18 EMAIL2 20 EMAIL CONFIGURATION 23 Configuration Data 26 Warranty Information 27 Tables Table 1 STATUS screen functional descriptions 9 Table 2 EDIT screen funct...

Page 4: ...rea network connection Power to the IntelliSwitch should be OFF Connect the IWS to the network via an RJ 45 network cable The RJ 45 connection on the IntelliSwitch is located on the bottom of the Inte...

Page 5: ...2 Figure 1 In the orange column on the left locate the word Network and click on it The image in Figure 2 should appear...

Page 6: ...3 Figure 2 Locate the word configuration at the top of the screen and click on it Figure 3 should appear...

Page 7: ...4 Figure 3 Update the fields on the configuration with the desired network settings and click Submit Make sure you cycle power on the unit before attempting to connect using the new network settings...

Page 8: ...me that is easily recognized for the device being used The IWS is now ready to use Refer to the section USING THE WEB SERVER to learn what functions it has and how to use it Figure 4 USING THE WEB SER...

Page 9: ...so it can be used Green In Use indicates that the left side is being used Red Replace indicates that the gas supply on the left side needs to be replaced 2B Cylinder Selection Radio buttons indicate...

Page 10: ...look back feature is invoked and after time out has determine that it can switch back to a side whose pressure has built back a switch back timer will start If the inlet pressure on the same side drop...

Page 11: ...gulator U P value Delivery Regulator Status value Reserve Status value Alarm Clear U P Left Side value Alarm Clear O P Left Side value Alarm Clear U P Right Side value Alarm Clear O P Right Side value...

Page 12: ...he Password pop up Figure 6 will appear Click in the blank text box and type 19490321 Now either click on OK or press ENTER on the keyboard The EDIT screen will appear Figure 7 Figure 6 Figure 7 The u...

Page 13: ...start If the inlet pressure on the same side drops again before the switchback time has expired the IntelliSwitch determines that that side is truly empty and will alarm and switch to the other inlet...

Page 14: ...n is selected asking the user to verify their intention to change the parameters 18 ZERO SET Button This button will zero the pressure displays It serves as a type of calibration feature If a pressure...

Page 15: ...or the current entry until the user clicks the mouse on another parameter or in the open space on the screen The method for changing all text boxes is the same Use the pop up message as a guide to ent...

Page 16: ...example it is initially set to HP or high pressure Figure 13 Using the Mouse click on the radio button that represents the selection to be made The radio button representing HP will change from green...

Page 17: ...ut will not have been transferred to the IntelliSwitch More modifications can be made or all the changes can be voided and the original values restored to the screen by clicking on the UNDO button or...

Page 18: ...U P under pressure Right Side U P value Reserve In Use Delivery Regulator O P value Delivery Regulator U P value Delivery Regulator Status value Alarm Clear U P Left Side value Alarm Clear O P Left S...

Page 19: ...f the person or system sending the email 5 SET button Clicking this button will save the parameters that have been entered 6 TEST button This button allows the user to test the email function Clicking...

Page 20: ...to report events An event is defined as any change that may occur to the IntelliSwitch Reported events include Inlet Value Status description Error description Left Side O P value O P over pressure Ri...

Page 21: ...f the person or system sending the email 5 SET button Clicking this button will save the parameters that have been entered 6 TEST button This button allows the user to test the email function Clicking...

Page 22: ...ondary DNS Servers Default Gateway Overriding Domain and Server Port may need to be identified and entered These values are set in the System Configuration screen which is accessed by clicking on the...

Page 23: ...can be accessed It is only shown for reference and cannot be changed on this screen To change the IP address only if needed it is necessary to use the Device Installer software that was included in yo...

Page 24: ...t number on that server Most services have a standard port 25 This port is used in conjunction with the Overriding Domain 7 SET button Clicking this button will save any new values entered into the sc...

Page 25: ...ove the header footer sections of your web browser F11 toggles full screen ON and OFF Press again to return to normal screen Web server is not loading on the screen No power to the IntelliSwitch Ether...

Page 26: ...___ Primary DNS Server ______________________________________ Secondary DNS Server ____________________________________ Default Gateway _________________________________________ Overriding Domain ____...

Page 27: ...other than CONTROLS CORPORATION OF AMERICA or if the equipment has been subject to abuse misuse negligence or accident CONTROLS CORPORATION OF AMERICA s sole and exclusive obligation and Buyer s sole...

Page 28: ...ADI 9513 D Certified ISO 9001 2000 Controls Corporation of America 1501 Harpers Road Virginia Beach VA 23454 Telephone 1 800 225 0473 or 757 422 8330 Fax 757 422 3125 www concoa com...

Reviews: