SAFETY
THIS PRODUCT IS NOT INTENDED FOR USE WITH TOXIC GASES, FLAMMABLE
GASES OR CORROSIVE GASES.
NOTICE
The Medical
IntelliSwitch™ is intended for use in medical installations. This product
meets NFPA-99 requirements for switchover systems. Compliance with NFPA-99 for
a complete installation (i.e. Cylinders, piping, alarms, etc.) is the responsibility of the
installer. Refer to NFPA-99 for complete system requirements.
Basic safety precautions must be followed to reduce the risk of fire, electrical shock or injury.
Connect the Medical IntelliSwitch™ to the correct line voltage. A label on the product
identifies what voltage it is wired for. CONNECTION TO AN INCORRECT VOLTAGE
CAN CAUSE SERIOUS DAMAGE TO THE PRODUCT AND WILL VOID ANY
WARRANTY.
While the Medical IntelliSwitch™ Switchover is dust and moisture resistant, it is NOT
water-proof or completely sealed. It should be installed where it will not be subjected to rain
or high concentrations of dust. Never pour or spray liquids directly onto the product.
Install the Medical IntelliSwitch™ where the ambient temperature range is between 0° F and
140° F.
Do not install this product in a hazardous environment.
If product appears damaged in any way, do not use and request service from CONCOA.
Consult the cylinder distributor for the proper use of cylinders and for any restrictions on
their use (such as flow rate and temperature requirements).
Store cylinders with valve caps screwed on, and chain cylinders to a supporting wall or
column.
Handle cylinders carefully and only with valve caps screwed on. The cap will reduce the
chance that the cylinder valve will break off if the cylinder is accidentally dropped or falls
over. The cap also protects the cylinder valve from damage to screw threads, which could
cause leaky connections.
No smoking should be permitted near oxygen, nitrous oxide, any other oxidizer, flammable
gases, or flammable mixtures, or in areas where cylinders are stored.
Summary of Contents for Medical IntelliSwitch 570 Series
Page 3: ...Blank Page ...
Page 9: ...7 Figure 1a ...
Page 10: ...8 Figure 1b ...
Page 16: ...14 Figure 2 ...
Page 27: ...25 Figure 12 ...
Page 44: ...42 MOUNTING DETAIL Figure 17 ...
Page 46: ...44 SELF TEST WORKSHEET Page 1 ...
Page 47: ...45 SELF TEST WORKSHEET Page 2 ...
Page 50: ...48 Blank Page ...