Test 5 - Display Switchover Pressure
The number on the display is the switchover
(changeover) pressure value that is internally
set for product operation. The value is in the
units of measure selected for the product (PSI,
BAR, etc.)
Test 6 – Auxiliary Settings
Displays settings for Units of Measure,
Communication setting, Keypad lockout,
allow remote settings enable, Reserve Gas
source enable.
Refer to Table 9 to determine what is enabled
Test 7 – Delivery Regulator Pressure
Tolerance
This number represents the value added or
subtracted to the nominal delivery pressure of
the product. It is used to determine the
allowable range of operation before a delivery
regulator fault condition is declared.
Test 8 – Delivery Regulator Alarm Delay
This value is the number of minutes the
system waits before looking back.