23
Using the pin assignments shown in Table 6 and the connector views in Figure 10 and Figure 11,
solder the wires to the appropriate pins. After soldering is complete, slide the locking nut over the
connector body. Snap the cable retainer into the body and then slide the protective cover over the
cable retainer.
Table 6
— Remote Alarm Pin Assignments
Medical
IntelliSwitch™
Alarm Output Connectors (A or B)
Pin Numbers
5740001
Alarm
Pin Number
Function
1
1
Left Inlet Alarm
2
2
Delivery Regulator Alarm.
3
3
+12v (provided by the remote
alarm)
4
4
Right Inlet Alarm
5
5
Reserve Gas Supply Alarm.
6
6
Ground (provided by the remote
alarm)
7
7
N.C.
8
8
Reserve In-Use
Wire terminations to the Alarm side of the cable are done in the same fashion as the Medical
IntelliSwitch™ connector.
Slide the protective cover, cable retainer, and locking nut over the end of the cable as shown in
Figure 10.