18
6.
After the pressure is set to the desired value, press the Output Source Pushbutton on the
front panel to switch to the other delivery regulator.
7.
Place the Allen wrench into the set screw of the second regulator and adjust the pressure to
the desired value.
It is possible to switch back to the other delivery regulator and change its pressure by
pushing the Output Source PB again. The delivery pressure for both regulators should be
the same to within 1-2 PSI for best performance.
8.
When the desired delivery pressure for both regulators is set, turn OFF SW4-4 on the control
board. See Figure 4.
9.
Locate SW5 on the control board (Figure 4) and press it once. The system should go back to
normal operation showing inlet pressures on both inlet displays. If an “Err2” message
appears when SW% is pressed it means that both regulators were not selected for calibration
(See step 6).
10.
Calibration is complete. Close door and re-attach screws.