13
Testing
GAS WILL CONTINUE TO FLOW WHILE IN THE TEST MODE.
Holding the Reset button in for 5 seconds will place the system in test mode. This diagnostic and
troubleshooting tool performs 23 different tests. Each time the Reset button is pushed again, the
system will perform a different test. Continuing to press the Reset button will step through all the
tests after which the system will return to normal operation. If after the test mode has been
entered there is no pushbutton activity for 5 minutes, the system will return to the normal display
operation. Gas flow is maintained during the test mode. Refer to the Troubleshooting Section for
details on testing.
Keypad Lockout
If the keypad security lockout feature is enabled, the keypad will become inoperative after
approximately 5 minutes provided there has been no buttons pressed. There will be an audible beep
when the keypad goes into the locked state. If the keypad becomes locked it can be re-activated by
holding the Bank Select Pushbutton for approximately 4 seconds. The system will make an audible
beep when it is unlocked. To enable this feature, see Section “User Selectable Features” for
explanation on enabling/disabling this feature.
USER SELECTABLE FEATURES
The Medical IntelliSwitch™ allows various system parameters to be set in the field. They are:
a.
Switchover Pressure
b.
Delivery Regulator Pressure
c.
Delivery Regulator Pressure Tolerance
d.
Units of Measure
e.
Keypad Security Lockout
f.
Remote Alarm Configuration
g.
Enabling Reserve Gas Supply
h.
Enabling Remote Setup
i.
Communication Port
Summary of Contents for Medical IntelliSwitch 570 Series
Page 3: ...Blank Page ...
Page 9: ...7 Figure 1a ...
Page 10: ...8 Figure 1b ...
Page 16: ...14 Figure 2 ...
Page 27: ...25 Figure 12 ...
Page 44: ...42 MOUNTING DETAIL Figure 17 ...
Page 46: ...44 SELF TEST WORKSHEET Page 1 ...
Page 47: ...45 SELF TEST WORKSHEET Page 2 ...
Page 50: ...48 Blank Page ...