MOUNTING REQUIREMENTS
Figure 3
INSTALLATIONS INSTRUCTIONS
Included with the Advantium 1
Plus
Alarm are four (4) screws and wall anchors. If using
the wall anchors drill a 1/8” hole in each of the four locations of the mounting holes of
the alarm. The alarm enclosure or the template provided may be used. Place the wall
anchors into the wall and attach the alarm using the screws provided.
Attach the output cable from the equipment being monitored to the Input of the alarm. If
using the output of the alarm to connect to another alarm or system, attach the output
cable at this time.
Plug the alarm into the appropriate power. Turn alarm ON by turning the Power switch
on the bottom of the alarm to the ON position and test the system. The system may be
tested once it is connected to the equipment being monitored by activating the switch on
the equipment. In the case of a CONCOA protocol system dropping the inlet pressure on
the system to below the trip point set on the pressure gauge will activate the alarm. Refer
to the operator’s manual for the protocol system to learn how to set the trip point.