14
3: Installation
3.1 Site Requirements
The
Dual O2/CO2
monitor enclosure should be mounted in an area free of vibration and
electrical noise or interference. If possible, avoid areas with high temperatures or
condensing humidity.
WARNING:
The
Dual O2/CO2
monitor is not designed for installation in hazardous
areas. Consult CONCOA for information on enclosures for use in hazardous
environments.
3.2 Mounting
3.2.1 Transmitter Enclosure
The
Dual O2/CO2
monitor is designed primarily for wall mounting and should be
installed at a height convenient for operation, maintenance, and viewing of the
instrument display. The following is a drawing of the mounting dimensions.
3.2.2
Dual O2 /CO2
monitor
The transmitter and sensor should be installed in a location where gas leaks are likely to
occur or where released gases may accumulate. It should be mounted no closer than
12 inches above floor level. Airflow within the monitored area, the characteristics of the
gas (lighter or heavier than air), and the position of workstations and personnel should
all be considered in determining the most suitable installation location.
6.5”